Summary

Eligibility
for people ages 18-49 (full criteria)
Healthy Volunteers
healthy people welcome
Location
at San Francisco, California
Dates
study started
completion around
Principal Investigator
by Philip Bickler, MD, PhD
Headshot of Philip Bickler
Philip Bickler

Description

Summary

The goal of this study was to evaluate the accuracy of pulse oximeters over the range of 70-100% per the International Organization for Standardization (ISO) 80601-2-61:2019. Four test devices were placed on each subject on the left or right hand with two at the finger base and two at the fingertip. Motion equipment was used to control motion of the test devices. In addition, two test devices were placed on the contralateral hand on the base of a finger and also on the fingertip. Finally, a test wrist device was placed on either the left or right wrist. SpO2 measurements from these devices were compared to sampling of arterial blood during brief stable oxygen desaturation in healthy volunteers to evaluate the claimed range.

Official Title

Accuracy of the Evie Ring Pulse Oximeter in Profound Hypoxia

Details

Keywords

Hypoxia, Pulse oximeter

Eligibility

Location

  • Hypoxia Research Laboratory accepting new patients
    San Francisco California 94133 United States

Lead Scientist at UCSF

  • Philip Bickler, MD, PhD
    Professor, Anesthesia, School of Medicine. Authored (or co-authored) 140 research publications

Details

Status
accepting new patients
Start Date
Completion Date
(estimated)
Sponsor
Movano Health
ID
NCT06459206
Study Type
Observational [Patient Registry]
Participants
Expecting 12 study participants
Last Updated